Changing Your Max Basal Amount
Device: 523/723 (Revel), 522/722, 515/715, 512/712
Important
Determine your Max basal rates with your healthcare provider.
| Allowed Range | Default Setting | Increments | |
|---|---|---|---|
| 523/723 (Revel) | 0.0-35.0 units/hour | 2.0 units/hour |
0.025 units (for rates between 0.025-0.975 u/h) 0.05 units (for rates between 1-9.95 u/h) 0.1 units (for rates of 10 u/h or more) |
| 522/722 | 0.0-35.0 units/hour | 2.0 units/hour | 0.05 units |
| 515/715 | 0.0-35.0 units/hour | 2.0 units/hour | 0.05 units |
| 512/712 | 0.0-35.0 units/hour | 2.0 units/hour | 0.05 units |
Go to the MAX BASAL RATE screen. The Max basal rate will be flashing.
For 523/723 (Revel)
Main Menu > Basal > Basal Setup > Max Basal RateFor 522/722, 515/715, 512/712 insulin pump models
Main Menu > Basal > Max Basal RateChange the rate and press ACT.
Your Max basal rate is now set. Exit the menus.