Setting Your Sensor Settings
Setting the Glucose Alert Limits
Device: 523/723 (Revel), 522/722, Guardian
Did You Know
The HIGH Sensor Glucose (SG) and LOW SG alerts notify you when your glucose reaches or goes above or below your programmed glucose limits. For example, if your LOW SG alert is set at 70 mg/dL, you will receive an alert every time your glucose reaches or goes below 70.
Important
Your glucose limits are not the same as your target glucose ranges. Work with your healthcare provider to determine your glucose alert limits.
| Allowed Low Range | Allowed High Range | Default Low Setting | Default High Setting | |
|---|---|---|---|---|
| 523/723 (Revel) |
40–390 mg/dL (2.2–21.6 mmol/L) | 50–400 mg/dL (2.8–22.2 mmol/L) | 80 mg/dL (4.4 mmol/l) | 240 mg/dL (13.3 mmol/L) |
| 523/723 K (Revel) |
90–390 mg/dL (5.0–21.6 mmol/L) | 100–400 mg/dL (5.6–22.2 mmol/L) | 80 mg/dL (4.4 mmol/l) | 280 mg/dL (15.5 mmol/L) |
| 522/722 | 40–390 mg/dL (2.2–21.6 mmol/L) | 110–400 mg/dL (2.8–22.2 mmol/L) | 50 mg/dL (2.8 mmol/L) | 200 mg/dL (11.2 mmol/L) |
| Guardian | 40–390 mg/dL (2.2 to 21.6 mmol/L) | 50–400 mg/dL (2.8–22.2 mmol/L) | 60 mgdL (3.3 mmol/L) | 200 mg/dL (11.2 mmol/L) |
| Guardian K | 90–390 mg/dL (5.0–21.6 mmol/L) | 100–400 mg/dL (5.6–22.2 mmol/L) | 90 mg/dL (5.0 mmol/L) | 280 mg/dL(15.5 mmol/L) |
Device: 523/723 (Revel), Guardian
Make sure the EDIT SETTINGS screen is open.
For 523/723 (Revel)
Main Menu > Sensor > Edit SettingsFor Guardian
Main Menu > Sensor > Sensor Setup > Edit Settings
Select Glucose Alerts (Glucose Alarms for Guardian), then press ACT. The GLUCOSE ALERTS screen appears.
Select On, then press ACT. The EDIT SETTINGS screen appears showing the glucose alerts are now turned on.
Make sure the EDIT SETTINGS screen is open.
Main Menu > Sensor> Edit Settings
Select Glucose Limits, then press ACT. The SET GLUCOSE LIMITS 1 screen appears. The Low Glucose Limit flashes. The start time for your first pair of Glucose Limits is midnight and cannot be changed.
Select your Low Glucose Limit.
Press ACT to select your High Glucose Limit. Your High Glucose Limit must also be at least 10 mg/dL (0.6 mmol/L) above your Low Glucose Limit.
Press ACT. The SET START TIME 2 screen appears. The dashes under the screen name flash. The first pair of Glucose Limits is now set.
If you do not need a second pair of Glucose Limits, press ESC. If you want to set up a second pair of Glucose Limits, follow steps 10 through 15 below.
To set a second pair of Glucose Limits, select the time of day you want this pair to become active.
Press ACT. The SET GLUCOSE LIMITS 2 screen appears. The default Low Glucose Limit flashes (OFF).
Select your Low Glucose Limit.
Press ACT. The default High Glucose Limit flashes (OFF).
Select your High Glucose Limit.
Press ACT. The SET START TIME 3 screen appears.
The second pair of Glucose Limits (Glucose Limits 2) is now set. If you do not need any more Glucose Limits pairs, press ESC. If you want to set more Glucose Limits, repeat steps 10–15, up to a maximum of 8.
Device: 522/722
Low Glucose Alert
Make sure the EDIT SETTINGS screen is open.
Main Menu > Sensor > Sensor Setup > Edit Settings
Select Low Glucose, then press ACT.
Select On, then press ACT. The SET LOW GLUCOSE screen flashes the default Low Glucose limit.
Select your Low Glucose limit. Your Low Glucose limit must be at least 10 mg/dL (0.6 mmol/L) below your High Glucose limit.
Press ACT. The EDIT SETTINGS screen shows the Low Glucose value set.
High Glucose Alert
Make sure the EDIT SETTINGS screen is open.
Main Menu > Sensor > Sensor Setup > Edit Settings
Select High Glucose, then press ACT.
Select On, then press ACT. The SET HIGH GLUCOSE screen flashes the default High Glucose.
Select your High Glucose limit. Your High Glucose limit must be at least 10 mg/dL (0.6 mmol/L) above your Low Glucose limit.
Press ACT. The EDIT SETTINGS screen shows the High Glucose value you selected.
Setting the Predictive Glucose Alert
Device: 523/723 (Revel), Guardian
Did You Know
Predictive alerts provide you with a warning that your glucose is trending towards your set high and low glucose limits 5–30 minutes before you reach these limits. With predictive alerts, you can be made aware of potential highs and lows even before they occur.
Important
Set up predictive alerts as recommended by your healthcare provider.
| Allowed Range | Default Setting | Increments | |
|---|---|---|---|
| 523/723 (Revel) | 5–30 minutes | 15 minutes | 5–30 minutes |
| Guardian | 5–30 minutes | 15 minutes | 5–30 minutes |
Make sure the EDIT SETTINGS screen is open.
For 523/723 (Revel)
Main Menu > Sensor > Edit SettingsFor Guardian
Main Menu > Sensor > Sensor Setup > Edit Settings
Select Predict Alerts (Predictive Alm for Guardian), then press ACT. The PREDICTIVE ALERTS LOW/HIGH screen shows that the alerts are turned Off.
Select On, then press ACT. The SET TIME SENSITIVITY LOW/HIGH screen appears and the LOW time setting flashes the default time.
Select the Time Sensitivity (how many minutes before you reach your Low Glucose Limit). To turn off the Low Predictive Alert, reduce the Time Sensitivity to OFF.
Press ACT. The HIGH time setting flashes the default time.
Select the Time Sensitivity (how many minutes before you reach your High Glucose Limit). To turn off the High Predictive Alert, reduce the Time Sensitivity to OFF.
Press ACT. The EDIT SETTINGS screen appears.
Setting the Rate of Change Alerts
Device: 523/723 (Revel), Guardian
Did You Know
Rate of change alerts notify you that your glucose levels are rising or falling at a certain rate. The RISE RATE alert can be helpful if you are prone to forgetting to take insulin or miscalculate a dose since glucose levels tend to spike significantly at these times. Likewise, the FALL RATE alert can be helpful in protecting you when you give yourself too much insulin or do not eat enough food.
| Allowed Range | Default Setting | Increments | |
|---|---|---|---|
| 523/723 (Revel) | 1.1–5.0 mg/dL/min (0.065 0.275 mmol/L/min) | Off | 0.1 mg/dL/min (0.005 mmol/L/min) |
| Guardian | 1.1–5.0 mg/dL/min (0.061 to 0.277 mmol/L/min) | Off | 0.1 mg/dL/min (0.005 mmol/L/min) |
Make sure the EDIT SETTINGS screen is open.
For 523/723 (Revel)
Main Menu > Sensor > Edit SettingsFor Guardian
Main Menu > Sensor > Sensor Setup > Edit Settings
Select Rate Alerts (Rate Alarms for Guardian), then press ACT. The SET FALL RATE LIMIT screen flashes the rate alert default setting: OFF.
Select the Fall Rate Limit.
Press ACT. The SET RISE RATE LIMIT screen flashes the rate alert default setting: OFF.
Select the Rise Rate Limit. The range and increments are the same as the Fall Rate Limit.
Press ACT. The Rate alerts will now use the settings you specified.
Note
You will be asked first to set a fall rate of 1.1 to 5.0 mg/dL/min, or select to turn it Off. After your fall rate is selected, you will then be asked to set your rise rate. You may set a rise rate of 1.1 to 5.0 mg/dL/min, or turn it Off. It's often best to wait until you have adjusted to CGM to set these alerts.